Open an image within a project, view a SIC or CLIC image, and then File > Save as to a new image file caused an error. The same error occurred for any image with an amount table set.
Workaround: Copy the image file to a location outside of the project > Open Image > Quantification > Amount Table > Clear Table > OK. Then, a SIC or CLIC image can be saved without problems.
Method > Run a script with the Baseline Correction command, and then View > Journal. The journal showed current configuration settings instead of the parameters in the script.
When baseline correction or undo is performed on MS data, Cancel button was enabled on the progress dialog although the operation cannot be cancelled.
Edit > Local Blob Detection within a tiny selected rectangle might fail with uncaught exception if the smoothing factors are set too big.
Quantification > Calibration Table, and click on Calibrant column to sort the calibrants. The order appeared to be random instead of by calibrant name.
Method > Run Script with a saved Local Blob Detection command might detect blobs at a wrong location if some graphic object was drawn before the script is run.
If a project had a run date file that has "&" in its file name and is associated with a vial, the project was not saved correctly. Trying opening the saved project will cause errors and missing vials.
Workaround: Manually remove "&" from a run data file name before it is imported into a project.
Processing runs associated with a vial that has ".csv" or ".xml" at the end of its name would fail with errors.
Workaround: Avoid naming vials with either of the problem extensions.
An analysis created by v2.4 could not be saved after additional images were loaded using newer versions.
Workaround: If source images are still available, open the analysis and File > Reload All before saving.
Open an existing analysis, load a new image, and then File > Save Analysis to the same analysis (GCA) folder. Try to load another new image, the program might report either the template or configuration file missing.
Workaround: Choose a different location to save an existing analysis that has template and/or configuration files.
Loading additional images to an open analysis without a specified feature template might use a recently used template previously applied to another analysis by mistake.
Analysis > Multi-class Statistics > Summary perspective, set some Scatter Chart for Compound Sets , then switching between charts and views might cause an exception.